EasyManua.ls Logo

VTI Instruments EX1200 Series - EX1200-6101 Specifications

VTI Instruments EX1200 Series
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.vtiinstruments.com
EX1200-6101 Plug-in Module 123
EX1200-6101
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
10 SP4T multiplexers
R
ELAY
T
YPE
Electromechanical, fail-safe
M
AXIMUM
S
WITCHING
V
OLTAGE
220 V dc, 250 V ac rms
M
AXIMUM
S
WITCHING
C
URRENT
2 A
M
AXIMUM
S
WITCHING
P
OWER
50 W, 62.5 VA
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
5 x 10
6
1 x 10
5
at full load
S
WITCHING
T
IME
< 5 ms
P
ATH
R
ESISTANCE
< 0.250 Ω
I
NSULATION
R
ESISTANCE
> 1 x 10
9
Ω
B
ANDWIDTH
(-3
dB)
1.3 GHz (typical)
I
NSERTION
L
OSS
(T
YPICAL
)
500 MHz
1.3 GHz
< 0.9 dB
< 3.0 dB
C
ROSSTALK
(T
YPICAL
)
500 MHz
1.3 GHz
< -70 dB
< -60 dB
I
SOLATION
(T
YPICAL
)
500 MHz
1.3 GHz
< -70 dB
< -60 dB
VSWR
(T
YPICAL
)
500 MHz
1.3 GHz
< 1.11:1
< 2.92:1
For mating connector, crimp pins, and other accessories, please refer to Appendix B.