VTI Instruments Corp.
42 EX1200-3001DS Plug-in Module
EX1200-3001DS
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
Eight individual (1 x 8) 2-wire or eight (1 x 16) 1-wire
R
ELAY
T
YPE
Electromechanical, fail-safe
M
AXIMUM
S
WITCHING
V
OLTAGE
300 V ac rms, 300V dc
M
AXIMUM
S
WITCHING
C
URRENT
2 A
M
AXIMUM
S
WITCHING
P
OWER
60 W dc, 125 VA
*Maximum switched power is at 30 V/ 2 A dc. Max switched power is derated non-linearly as voltage is increased.
M
INIMUM
C
ONTACT
R
ATING
*
10 mV dc, 10 µA (resistive)
*This value is in reference to a resistive load. Minimum capacity changes depending on switching frequency and environmental conditions
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
1 x 10
(no load)
1 x 10
6
@ 50 V dc, 0.1 A resistive or 10 V dc, 10 mA (resistive)
S
WITCHING
T
IME
< 3 ms
P
ATH
R
ESISTANCE
< 0.5 Ω
I
NSULATION
R
ESISTANCE
> 1 x 10
Ω
M
AXIMUM
T
HERMAL
O
FFSET
PER
C
HANNEL
(HI-LO)
< 1 μV
C
APACITANCE
Open channel
Channel-mainframe
High-low
< 50 pF
< 20 pF
< 50 pF
B
ANDWIDTH
(-3
dB)
50 MHz (typical)
I
NSERTION
L
OSS
100 kHz
1 MHz
10 MHz
< 0.1 dB
< 0.2 dB
< 0.5 dB
C
ROSSTALK
100 kHz
1 MHz
10 MHz
< -90 dB
< -70 dB
< -50 dB
I
SOLATION
100 kHz
1 MHz
10 MHz
< -90 dB
< -70 dB
< -60 dB
For mating connector, crimp pins, and other accessories, please refer to Appendix B.