VTI Instruments Corp.
88 EX1200-4128Plug-in Module
EX1200-4128
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
4 x 128 one-wire cross-point matrix
R
ELAY
T
YPE
Electromechanical, fail-safe
M
AXIMUM
S
WITCHING
V
OLTAGE
250 V AC, 220 V DC
M
AXIMUM
S
WITCHING
C
URRENT
1 A
M
AXIMUM
S
WITCHING
P
OWER
60 W dc, 62.5 VA
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
1 x 10
8
1 x 10
5
at full load
S
WITCHING
T
IME
5 ms typical
P
ATH
R
ESISTANCE
< 1 Ω
I
NSULATION
R
ESISTANCE
> 1 x 10
9
Ω
M
AXIMUM
T
HERMAL
O
FFSET
PER
C
HANNEL
(HI-LO)
< 10 µV
B
ANDWIDTH
(-3
dB)
3 MHz (typical)
C
ROSSTALK
1 MHz
10 MHz
< -55 dB< -30 dB
I
SOLATION
1 MHz
10 MHz
< -60 dB
< -30 dB
For mating connector, crimp pins, and other accessories, please refer to Appendix B.