www.vtiinstruments.com
EX1200-7100 Plug-in Module 151
EX1200-7100
S
PECIFICATIONS
RF
P
ERFORMANCE
S
PECIFICATIONS
Frequency (GHz)
VSWR
Insertion loss (dB max)
Isolation (dB min)
RF power (CW)
DC – 4 4 – 8 8 – 12 12 – 18 18 – 26
1.25:1 1.35:1 1.40:1 1.50:1 1.8:1
0.20 0.30 0.40 0.50 0.80
70 65 60 60 50
100 70 60 45 30
A
DDITIONAL
S
PECIFICATIONS
Maximum operating voltage
Maximum current rating (carrier)
Maximum current per switch
7102/7104/7106
7122
Switching time
Operating mode
7102/7104
7106
7122
Switching sequence
Operating life
24 V dc
1 A
140 mA @ nominal voltage and 25 °C
280 mA @ nominal voltage and 25 °C
20 ms (maximum)
Failsafe
Normally open
Transfer, failsafe
Break-Before-Make
2,000,000 cycles per position minimum