EasyManua.ls Logo

VTI Instruments EX1200 Series - EX1200-2007 A Specifications

VTI Instruments EX1200 Series
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.vtiinstruments.com
EX1200-2007A Plug-in Module 29
EX1200-2007A
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
(1 x 24) 2-wire, dual (1 x 12) 2-wire, or dual (1 x 24) 1-wire
R
ELAY
T
YPE
Reed
M
AXIMUM
S
WITCHING
V
OLTAGE
1000 V dc / 700 V ac rms
M
AXIMUM
S
WITCHING
C
URRENT
1 A
M
AXIMUM
C
ARRY
C
URRENT
2 A
M
AXIMUM
S
WITCHING
P
OWER
25 W (resistive load)
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
1 x 10
8
1 x 10
6
(full load)
S
WITCHING
T
IME
< 1 ms
P
ATH
R
ESISTANCE
< 1 Ω
I
NSULATION
R
ESISTANCE
> 1 x 10
7
Ω
B
ANDWIDTH
(-3
dB)
65 MHz (typical)
For mating connector, crimp pins, and other accessories, please refer to Appendix B.
Application Note
The EX1200-2007A is intended to be used as a multiplexer only and not as a splitter (i.e. split a single input signal
into multiple paths). Use of the EX1200-2007A as a splitter may cause damage to its circuitry.

Table of Contents