EX1200-7100 Plug-in Module 159
EX1200-08442
S
PECIFICATIONS
G
ENERAL
S
PECIFICATIONS
C
HANNEL
C
OUNT
44*8 2wire
R
ELAY
T
YPE
Electromechanical, fail-safe
M
AXIMUM
S
WITCHING
V
OLTAGE
300 V dc, 300 V ac rms
M
AXIMUM
S
WITCHING
C
URRENT
2 A
M
AXIMUM
S
WITCHING
P
OWER
60 W dc, 125 VA
*Maximum switched power is at 30 V/ 2 A dc. Max switched power is derated non-linearly as voltage is increased.
M
INIMUM
C
ONTACT
R
ATING
*
10 mV dc, 10 µA (resistive)
*This value is in reference to a resistive load. Minimum capacity changes depending on switching frequency and environmental conditions
R
ATED
S
WITCH
O
PERATIONS
Mechanical
Electrical
1 x 10
8
(no load)
1 x 10
6
@ 50 V dc, 0.1 A resistive or 10 V dc, 10 mA (resistive)
S
WITCHING
T
IME
< 3 ms
P
ATH
R
ESISTANCE
< 366 mΩ for one Slot
< 630 mΩ between SLOT1 and SLOT2
< 950 mΩ between SLOT1 and SLOT4
I
NSULATION
R
ESISTANCE
> 1 x 10
9
Ω
M
AXIMUM
T
HERMAL
O
FFSET
PER
C
HANNEL
(HI-LO
)
<1 μV
C
APACITANCE
High-low
<240 pF
B
ANDWIDTH
(-3
dB)
30 MHz for SINGLE SLOT (Typical)
21 MHz between SLOT1 and SLOT2 (Typical)
15 MHz between SLOT1 and SLOT4 (Typical)
C
ROSSTALK
(T
YPICAL
)
250 KHz
1 MHz
< -26 dB
< -50 dB
I
SOLATION
(T
YPICAL
)
250 KHz
1 MHz
< -28 dB
< -52 dB