VTI Instruments Corp.
56 EX1200-3072 Plug-in Module
TB
Ref Signal
Conn
Pin
TB
Ref Signal
Conn
Pin
TB
Ref Signal
Conn
Pin
TB
Ref Signal
Conn
Pin
T1
CH1_1L E2
T41
CH1_21L C10
T81
CH2_1L A18
T121
CH2_21L E26
T2
CH1_1H E3
T42
CH1_21H C11
T82
CH2_1H A19
T122
CH2_21H E27
T3
CH1_2L C4
T43
CH1_22L E10
T83
CH2_2L D18
T123
CH2_22L E24
T4
CH1_2H C5
T44
CH1_22H E11
T84
CH2_2H D19
T124
CH2_22H E25
T5
CH1_3L A2
T45
CH1_23L D12
T85
CH2_3L C18
T125
CH2_23L A24
T6
CH1_3H A3
T46
CH1_23H D13
T86
CH2_3H C19
T126
CH2_23H A25
T7
CH1_4L B2
T47
CH1_24L A10
T87
CH2_4L D16
T127
CH2_24L D28
T8
CH1_4H B3
T48
CH1_24H A11
T88
CH2_4H D17
T128
CH2_24H D29
T9
CH1_5L A4
T49
CH1_25L D10
T89
CH2_5L C16
T129
CH2_25L B28
T10
CH1_5H A5
T50
CH1_25H D11
T90
CH2_5H C17
T130
CH2_25H B29
T11
CH1_6L B4
T51
CH1_26L B8
T91
CH2_6L D20
T131
CH2_26L C26
T12
CH1_6H B5
T52
CH1_26H B9
T92
CH2_6H D21
T132
CH2_26H C27
T13
CH1_7L D2
T53
CH1_27L D8
T93
CH2_7L A20
T133
CH2_27L A26
T14
CH1_7H D3
T54
CH1_27H D9
T94
CH2_7H A21
T134
CH2_27H A27
T15
CH1_8L C2
T55
CH1_28L B14
T95
CH2_8L B18
T135
CH2_28L C30
T16
CH1_8H C3
T56
CH1_28H B15
T96
CH2_8H B19
T136
CH2_28H C31
T17
CH1_9L D4
T57
CH1_29L C14
T97
CH2_9L E18
T137
CH2_29L D26
T18
CH1_9H D5
T58
CH1_29H C15
T98
CH2_9H E19
T138
CH2_29H D27
T19
CH1_10L A6
T59
CH1_30L C12
T99
CH2_10L D22
T139
CH2_30L A28
T20
CH1_10H A7
T60
CH1_30H C13
T100
CH2_10H D23
T140
CH2_30H A29
T21
CH1_11L E4
T61
CH1_31L E12
T101
CH2_11L A22
T141
CH2_31L A30
T22
CH1_11H E5
T62
CH1_31H E13
T102
CH2_11H A23
T142
CH2_31H A31
T23
CH1_12L C8
T63
CH1_32L B16
T103
CH2_12L B22
T143
CH2_32L E28
T24
CH1_12H C9
T64
CH1_32H B17
T104
CH2_12H B23
T144
CH2_32H E29
T25
CH1_13L C6
T65
CH1_33L A16
T105
CH2_13L C20
T145
CH2_33L B30
T26
CH1_13H C7
T66
CH1_33H A17
T106
CH2_13H C21
T146
CH2_33H B31
T27
CH1_14L A8
T67
CH1_34L E14
T107
CH2_14L B20
T147
CH2_34L C28
T28
CH1_14H A9
T68
CH1_34H E15
T108
CH2_14H B21
T148
CH2_34H C29
T29
CH1_15L B6
T69
CH1_35L D14
T109
CH2_15L C24
T149
CH2_35L E30
T30
CH1_15H B7
T70
CH1_35H D15
T110
CH2_15H C25
T150
CH2_35H E31
T31
CH1_16L E8
T71
CH1_36L A12
T111
CH2_16L B24
T151
CH2_36L D30
T32
CH1_16H E9
T72
CH1_36H A13
T112
CH2_16H B25
T152
CH2_36H D31
T33
CH1_17L D6
T73
CH1_COML A14
T113
CH2_17L C22
T153
CH2_COML E16
T34
CH1_17H D7
T74
CH1_COMH A15
T114
CH2_17H C23
T154
CH2_COMH E17
T35
CH1_18L E6
T75
SHIELD E20
T115
CH2_18L E22
T155
SHIELD D32
T36
CH1_18H E7
T76
SHIELD E21
T116
CH2_18H E23
T156
SHIELD E32
T37
CH1_19L B10
T77
SHIELD E1
T117
CH2_19L D24
T157
SHIELD B32
T38
CH1_19H B11
T78
SHIELD D1
T118
CH2_19H D25
T158
SHIELD C32
T39
CH1_20L B12
T79
SHIELD C1
T119
CH2_20L B26
T159
GND_C A1
T40
CH1_20H B13
T80
SHIELD B1
T120
CH2_20H B27
T160
GND_C A32
T
ABLE
4-16:
EX1200-TB160-1
T
ERMINAL
B
LOCK TO
EX1200-3072
P
IN
M
APPING
RT1 can be measured by the EX1200 DMM or may be measured using an external instrument. If
an external sensor is used, it must be connected to T164 (L_VS) and T163 (H_VS). The user may
also choose to use a sensor other than the on-board thermistor. To do so, connect the sensor using
T162 (RL_I) and T161 (RH_I). Note that CH1_1L (T1) and CH1_1H (T2) must be dedicated to
making temperature measurements once P2 is configured.