EasyManua.ls Logo

Waters 2998 - Maintaining the Flow Cell

Waters 2998
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-4 Maintaining the Detector
Maintaining the flow cell
The flow cell requires maintenance when
the reference spectrum changes.
the cell fluid leaks out of the drain tube.
the detector cannot initialize but the lamp is in good condition.
the detector causes high backpressure.
Tip: Conditions other than a dirty flow cell can cause decreased lamp
intensity. For more information, refer to Chapter 4.
Flow cell maintenance consists of
flushing the flow cell.
removing the flow cell.
disassembling and cleaning the flow cell.
installing the flow cell assembly.
Flushing the flow cell
Required Materials
•HPLC-grade water
HPLC-grade methanol
If the flow cell requires cleaning, first try flushing it with solvent.
To flush the flow cell:
1. Select a solvent compatible with the samples and mobile phases that you
have been using. If you have been using buffers, flush with 10 mL of
HPLC-grade water, then flush with 10 mL of a low-surface-tension
solvent such as methanol.
Tip: Ensure that the solvent is miscible with the previous mobile phase.
2. Test the lamp energy by performing the Read energy diagnostic test (see
page 4-3).
If the lamp diagnostic test fails and the lamp has not been used more
than 2000 hours or 1 year from date of purchase (whichever comes first),
call Waters Technical Service (see page 3-2).

Table of Contents

Other manuals for Waters 2998

Related product manuals