EasyManua.ls Logo

Waters ACQUITY RDa - Page 22

Waters ACQUITY RDa
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 2–1: UNIFI setup and calibration -RDa
Tips:
If the required calibration is not available when setting up an analysis, the samples will not
run. If this happens, the software will guide you on how to complete the calibration.
When transitioning from Power Save mode to Operate mode, Waters recommend that you
allow the instrument to stabilize for one hour prior to use.
If the Start button on the instrument setup control panel is not available, the instrument is
currently not ready to run the setup.
To setup and create a calibration:
1. Click Setup.
2. Select at least one of the following calibration slots:
Positive low mass (50 - 2000 Da)
Positive high mass (400 - 7000 Da)
Negative low mass (50 - 2000 Da)
Negative high mass (400 - 5000 Da)
3. Click Start.
Notes:
A progress bar displays as the instrument is calibrated, to show how the calibration is
progressing.
The date and time of the successful calibration appears against the slot.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page 22

Table of Contents

Related product manuals