C Specifications
The applicability of the following specifications depends on the conditions in individual
laboratories. Refer to the ACQUITY RDa Detector System Site Preparation Guide, or contact the
Waters Technical Service organization for additional information about the specifications.
C.1 Physical specifications
The following table lists the physical specifications for the ACQUITY RDa detector.
Table C–1: Physical specifications
Attribute Standard RDa
Height 75.0 cm
Width 40.0 cm
Depth 75.0 cm
Weight 80 kg
C.2 Environmental specifications
The following table lists the environmental specifications for the ACQUITY RDa detector.
Table C–2: Environmental specifications
Attribute Specification
Operating temperature (performance is
specified)
15 °C to 28 °C
Operating humidity 20% to 80%, non-condensing
Temperature stability This must be better than 2 °C or 3.5 °F peak-to-
peak in 1.5 hours
Note: Fluctuations in temperature can affect the performance of the instrument. To avoid this,
ensure that the instrument operates within the specified ranges.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)