Cleaning the sampling cone assembly
January 11, 2016, 715004599 Rev. B 117
Disassembling the sampling cone assembly
Required material
• Chemical-resistant, powder-free gloves
• Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly:
1. Slide the collar to the end of the combined 2.5-mm Allen wrench and cone
extraction tool.
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated
surfaces, wear clean, chemical-resistant, powder-free gloves when working
with the sampling cone assembly.