4 Maintenance Procedures
174 January 11, 2016, 715004599 Rev. B
Replacing the IonSABRE II probe heater
Replace the IonSABRE II probe heater if it fails to heat the probe.
Removing the IonSABRE II probe heater
Required material
Chemical-resistant, powder-free gloves
To remove the IonSABRE II probe heater:
1. Remove the probe from the source (see page 64).
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves when working with the
probe and source components.