January 11, 2016, 715004599 Rev. B v
Safety hazard symbol notice
Documentation needs to be consulted in all cases where the symbol is used to find
out the nature of the potential hazard and any actions which have to be taken.
Considerations specific to the Xevo TQ-S micro
Power cord replacement hazard
Solvent leakage hazard
The source exhaust system is designed to be robust and leak-tight. Waters recommends
you perform a hazard analysis, assuming a maximum leak into the laboratory atmosphere
of 10% LC eluate.
Warning: To avoid electric shock, use the SVT-type power cord in the United
States and HAR-type (or better) in Europe. The main power cord must only be
replaced with one of adequate rating. For information regarding what cord to use in
other countries, contact your local Waters distributor.
Warning:
To confirm the integrity of the source exhaust system, renew the source
O-rings at intervals not exceeding one year.
To avoid chemical degradation of the source O-rings, which can
withstand exposure only to certain solvents (see page 258), determine
whether any solvents you use that are not listed are chemically
compatible with the composition of the O-rings.

Table of Contents

Other manuals for Waters Xevo TQ-S micro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters Xevo TQ-S micro and is the answer not in the manual?

Waters Xevo TQ-S micro Specifications

General IconGeneral
BrandWaters
ModelXevo TQ-S micro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals