4 Maintenance Procedures
172 January 11, 2016, 715004599 Rev. B
16. Fit the nebulizer adjuster knob to the PEEK union/UNF coupling assembly.
17. Finger-tighten the nebulizer adjuster knob onto the probe assembly.
18. Fit the probe gasket and end-cover to the probe assembly.
19. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit and tighten
the 3 screws retaining the probe end-cover.
20. Return the combined 2.5-mm Allen wrench and cone extraction tool in its storage
location on the source adaptor housing.
21. Fit the probe heater (see step 1 through step 3 on page 176).
22. Fit the probe to the instrument (see page 61).
23. In the Instrument Console, click API to start the probe and desolvation gas
flows.
Cleaning or replacing the corona pin
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
HPLC-grade (or better) methanol
Lint-free tissue
Lapping film
Notice:
To avoid damaging the probe heater, take great care to grip the heater so
as not to damage its electrical wiring.
To avoid damaging the probe heaters electrical connections, capillary
sleeve, or capillary, take great care when fitting the heater over the
capillary sleeve.

Table of Contents

Other manuals for Waters Xevo TQ-S micro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters Xevo TQ-S micro and is the answer not in the manual?

Waters Xevo TQ-S micro Specifications

General IconGeneral
BrandWaters
ModelXevo TQ-S micro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals