NanoFlow source
January 11, 2016, 715004599 Rev. B 73
The low flow rates involved with operating the NanoFlow source prohibit its use with the
instrument’s solvent delivery system.
Installing the NanoFlow source
Required material
Chemical-resistant, powder-free gloves
To install the NanoFlow source:
1. Prepare the instrument for working on the source (see page 97).
2. Remove the probe from the currently installed source.
If you are removing an ESI probe, see page 59.
If you are removing an IonSABRE II probe, see page 64.
3. Remove the existing source enclosure (see page 98).
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The source components can be contaminated.
Warning: To avoid electric shock, ensure that the instrument is prepared for
working on the source before commencing this procedure.
Warning: To avoid burn injuries, ensure that the source heater is turned off
and the block is cool before opening the source. The ion block, which can be
hot, is exposed when you fit the NanoFlow source.
Warning: To avoid burn injuries, take great care while working with the
probe and source; these components can be hot.

Table of Contents

Other manuals for Waters Xevo TQ-S micro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters Xevo TQ-S micro and is the answer not in the manual?

Waters Xevo TQ-S micro Specifications

General IconGeneral
BrandWaters
ModelXevo TQ-S micro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals