January 11, 2016, 715004599 Rev. B 261
D Plumbing the IntelliStart Fluidics
System
This appendix provides reference information for replacing the tubing in the
IntelliStart™ Fluidics system.
Warning: To avoid electric shock, do not use stainless steel tubing or
stainless steel “finger tight” screws to connect the selector valve to the
source probes; use the PEEK tubing and natural (beige) colored PEEK
“finger tight” screws specified on page 264.
Warning: To avoid personal contamination, always wear
chemical-resistant, powder-free gloves while working on the system.
Contents:
Topic Page
Preventing contamination............................................................................. 262
The selector valve......................................................................................... 262
Plumbing schematic...................................................................................... 263
Tubing and connection specifications........................................................... 264