Replacing the IonSABRE II probe heater
January 11, 2016, 715004599 Rev. B 175
2. Gripping the probe heater as shown, carefully pull it off the probe assembly.
3. Dispose of the probe heater in accordance with local environmental regulations.
Notice: To avoid damaging the probe heaters electrical connections, do not
twist the heater when removing it from the probe assembly.
Warning: To avoid spreading contamination, dispose of the probe
heater in accordance with local environmental regulations. The probe
heater can be contaminated with biohazardous or toxic materials.
Probe heater