Param
No.
EDS Values Read/Write
Default
Value
(EDS)
Parameter
Range
Comments
5 SetFailsafeEnable Write 0
0-1
(False/True)
Enabled the failsafe speed.
If disabled the pump will
stop in the event of a
communications loss. If
enabled the pump will run
at the speed set in the
SetFailsafeSpeed”
parameter
6 SetReverse Write 0
0-1
(False/True)
If set the pump will run
anti-clockwise. Pump
defaults to clockwise
rotation
7 Run Write 0
0-1
(False/True)
Set to 1 (true) to allow the
pump to run. 0 will stop
the pump. Note that
pump enable needs to be
set
8 RunEnable Write 0
0-1
(False/True)
Need to set to 1 to allow
pump to run. Setting to 0
will stop the pump and
not allow the pump to
run.
9 ResetRunHours Write 0
0-1
(False/True)
Resets the run hours
accumulator
10 PauseFlowTotaliser Write 0
0-1
(False/True)
Set to 1 to pause the
internal FlowTotaliser
parameter. Setting to 0
will un-pause the
parameter
11 ResetFlowTotaliser Write 0
0-1
(False/True)
Set to 1 to reset the flow
totaliser. Set to 0 to allow
the flow totaliser to
accumulate