22.8 Pump parameters
Setting parameters
To set a parameter to a new value:
• Type a value into the field or click on the check box (depending on the parameter type)
• Click 'set' to store the new value or click 'refresh' to cancel the change
• Up to 10 parameters are displayed per page. Use the < and > buttons to navigate between pages
The following parameters can be set by the user.
Param
No.
EDS Values Read/Write
Default
Value
(EDS)
Parameter
Range
Comments
1 SetFlowCal Write 15120
1 -
2147483647
Use to set the tube flow
calibration value. The flow
calibration value is set in
μL. For more information
on flow calibration
see"Flow calibration "
on page73
2 SetSpeed Write 100 1 - 2200
Speed is set in Deci RPM.
Max speed depends on
head type.See
"PumpHead" on
page108.
3 SetSpeedLimit Write 2200 1 - 2200
Speed is set in Deci RPM.
Max speed depends on
head type. See
"PumpHead" on
page108.
4 SetFailsafeSpeed Write 100 1 - 2200
If the failsafe is enabled
the pump will run
continuously at the
selected speed in the
event of a
communications loss.
m-630un-en-08 101