Servicing 11
4. Servicing
Important:
Remember: a functional check must be performed after every repair.
We recommend having all maintenance work, servicing and repairs carried out either by the manufacturer
Weinmann or by a qualified technician.
4.1 Intervals
The following checks must be carried out to ensure that the modules always function correctly.
Before each use:
• Carry out a final check (see sections 3.4 to
3.7).
After each use
• Clean, disinfect or sterilise the device and its
components (see “9. Hygienic preparation” in
the operating instructions).
• Carry out a final check (see sections 3.4 to
3.7).
At least every 6 months, if the equip-
ment has not been used in the interve-
ning period:
• Carry out a final check (see sections 3.2 to
3.4).
After every repair:
• Clean, disinfect or sterilise the device and its
components (see “9. Hygienic preparation” in
the operating instructions);
• Carry out a final check (see sections 3.4 to
3.7).
4.2 Servicing
The devices must be serviced every 4 years.
The following points must be covered:
• Visual inspection:
– mechanical damage;
– labelling of components.
• Pressure indicator accuracy;
• Freedom from leaks;
• Oxygen delivery rate and aspiration capacity;
• Renewal of parts subject to wear / mandatory
change parts (see “7.2 Spare parts required
for servicing” on page 37);
• Functional check in accordance with Test In-
structions / Test Record WM 22248 (see “3.
Final Check” on page 7 and see “11. Repair
and inspection log” on page 45 to 48).