Technical data 43
9.3 Aspiration efficiency
Maximal suction
The maximal suction exerted by the Combi and Suction MODULES depends on the input pressure of the
oxygen supply. The data below are valid for a 2 litre aspirate bottle.
Tolerance limits ±10%. If a bacterial filter is used,
the aspiration efficiency will diminish by between
1 and 3 l/min, depending on the atmospheric
humidity.
Aspiration efficiency of Suction MODULE
The suction exerted by the Suction MODULE varies according to the degree of vacuum selected at any given
time. The data below are valid for a 2 litre aspirate bottle.
Tolerance limits ±10%. If a bacterial filter is used,
the aspiration efficiency will diminish by between
1 and 3 l/min.
Input pressure p2 Max. vacuum Maximal suction Oxygen consumption
2.7 bar - 0.40 bar 8.4 l/min 9.3 l/min
3.0 bar - 0.44 bar 8.6 l/min 10.1 l/min
3.5 bar - 0.53 bar 8.8 l/min 11.4 l/min
4.0 bar - 0.58 bar 9.0 l/min 12.2 l/min
4.5 bar - 0.60 bar 9.0 l/min 14.0 l/min
5.0 bar - 0.60 bar 9.0 l/min 15.0 l/min
5.5 bar - 0.60 bar 8.7 l/min 17.0 l/min
6.0 bar - 0.60 bar 8.5 l/min 19.0 l/min
Selected vacuum Suction Oxygen consumption
Required input
pressure p2
- 0.2 bar 5.3 l/min 6.5 l/min 2.7 – 6 bar
- 0.3 bar 6.5 l/min 7.7 l/min 2.7 – 6 bar
- 0.4 bar 8.4 l/min 9.3 l/min 2.7 – 6 bar
- 0.5 bar 8.3 l/min 10.5 l/min 3.5 – 6 bar
- 0.6 bar 9.0 l/min 14.0 l/min 4.5 bar

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann Oxygen MODULE and is the answer not in the manual?

Weinmann Oxygen MODULE Specifications

General IconGeneral
BrandWeinmann
ModelOxygen MODULE
CategoryMedical Equipment
LanguageEnglish

Related product manuals