Loading...
Technical data 41
9. Technical data
Oxygen MODULE Suction MODULE Combi MODULE Interface MODULE
Product category
according to
93/42/EEC
IIa
Dimensions
L x B x H in mm
100 x 125 x 90 100 x 130 x 90 100 x 172 x 90
Weight 0.9 kg 0.9 kg 0.95 kg 0.65 kg
Temperature range
for
operation
storage
–18 °C to +60 °C
–40 °C to +70 °C
Gas input medicinal oxygen
Operating pressure 2.76 bar possible, 4.5bar recommended
Minimal gas volume
required.
80 l/min O
2
Pressure gas
connection
External thread G3/8; NIST adapter available on request
Oxygen flow
variable from
0 to 15 l/min
variable from
0 to 15 l/min
Tolerance limits
at 0 – 3.9 l/min
at 4 – 15 l/min
±15%
±10%
±15%
±10%
DIN ISO 10079-3
classification
medium vacuum