EasyManuals Logo

Weinmann VENTIlogic plus User Manual

Default Icon
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
Technical data 109
EN
Duration of battery charging
process
approx. 6 h
Classification to
EN 60601-1
type of protection against
electric shock
degree of protection against
electric shock
Protection class II
Type BF
Degrees of protection
Against the ingress of solid
particles
Against access to hazardous
parts
Against the harmful ingress of
water
IP21
Electromagnetic
compatibility (EMC) to
EN 60601-1-2:
radio interference suppression
radio interference immunity
EN 55011
EN 61000-3-2, EN 61000-3-3, EN 61000-4-2 to 6,
EN 61000-4-8, EN 61000-4-11
Mean sound pressure level/
operation to EN ISO 17510 at a
distance of 1 m from the device
in the patient position
approx. 28 dB (A) at 10 hPa
Sound pressure level of alarm
message
approx. 69 dB (A) to EN 60601-1-8
IPAP pressure range
EPAP / PEEP pressure range
CPAP pressure range
Pressure accuracy
Increment
6 to 40 hPa (leakage ventilation)
4 to 45 hPa (valve ventilation)
4 to 20 hPa (leakage ventilation)
0 to 20 hPa (valve ventilation)
4 to 20 hPa (leakage ventilation)
up to 35 hPa ± 0.8 hPa
from 35 hPa ± 1.5 hPa
0.2 hPa
(1 hPa = 1 mbar ö 1 cm H
2
O)
Minimum stable limit pressure
(PLS
min
) (min. pressure in the
event of a fault)
Maximum stable limit pressure
(PLS
max
) (max. pressure in the
event of a fault)
ø 0 hPa
÷ 60 hPa
Therapy device

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann VENTIlogic plus and is the answer not in the manual?

Weinmann VENTIlogic plus Specifications

General IconGeneral
BrandWeinmann
ModelVENTIlogic plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals