Duration of battery charging
process
approx. 6 h
Classification to
EN 60601-1
– type of protection against
electric shock
– degree of protection against
electric shock
Protection class II
Type BF
Degrees of protection
– Against the ingress of solid
particles
– Against access to hazardous
parts
– Against the harmful ingress of
water
IP21
Electromagnetic
compatibility (EMC) to
EN 60601-1-2:
– radio interference suppression
– radio interference immunity
EN 55011
EN 61000-3-2, EN 61000-3-3, EN 61000-4-2 to 6,
EN 61000-4-8, EN 61000-4-11
Mean sound pressure level/
operation to EN ISO 17510 at a
distance of 1 m from the device
in the patient position
approx. 28 dB (A) at 10 hPa
Sound pressure level of alarm
message
approx. 69 dB (A) to EN 60601-1-8
IPAP pressure range
EPAP / PEEP pressure range
CPAP pressure range
Pressure accuracy
Increment
6 to 40 hPa (leakage ventilation)
4 to 45 hPa (valve ventilation)
4 to 20 hPa (leakage ventilation)
0 to 20 hPa (valve ventilation)
4 to 20 hPa (leakage ventilation)
up to 35 hPa ± 0.8 hPa
from 35 hPa ± 1.5 hPa
0.2 hPa
(1 hPa = 1 mbar ö 1 cm H
2
O)
Minimum stable limit pressure
(PLS
min
) (min. pressure in the
event of a fault)
Maximum stable limit pressure
(PLS
max
) (max. pressure in the
event of a fault)
ø 0 hPa
÷ 60 hPa
Therapy device