B
97
EMC Guidance and Manufacturer’s
Declarations
Table 5. Electromagnetic Emissions
The CP 200 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 200
electrocardiograph should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The CP 200 electrocardiograph uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A The CP 200 electrocardiograph is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies