36 Chapter 3 Reviewing the System Settings Welch Allyn CP 200 Electrocardiograph
Configuring the Electrocardiograph for
Wireless Communication
Before you configure the electrocardiograph for wireless communication, verify the
following:
A wireless radio is installed in the electrocardiograph’s USB slot according to the
instructions that were shipped with the radio.
The CardioPerfect workstation software is installed according to the instructions that
were shipped with that CD.
The CardioPerfect workstation connectivity kit is installed according to the
instructions that were shipped with that CD.
To Configure the Electrocardiograph for Wireless Communication
1. At the electrocardiograph, press the Menu key .
2. Choose System Settings > Connectivity Settings > Connectivity Mode >
Wireless.
3. Under Network Settings, do one of the following:
Select the Use DHCP? checkbox. Press Done.
Enter the IP address, subnet mask, and gateway. Press Done.
4. Under Wireless Settings, fill in the fields — SSID, encryption, authentication,
network key, 802.1X auth, and EAP type — per your network administrator.
5. Test the connectivity settings. See
“Test Connectivity Settings” on page 32.