Technical Manual
SureTemp
®
Model 670/SureTemp 4
®
Model 675 21
Non-recoverable errors are generated from internal test failures that are catastrophic (or from promoted
recoverable errors). The error code and any appropriate icon will be displayed along with the “[X]”
Instrument Malfunction icon.
The only way to continue at this point is to reset the electronics by removing the batteries.
NOTE: The error codes E2.1, E2.3, E3.0, E3.1, E3.2 and E4.1 can sometimes be caused by a faulty probe.
It is advisable to remove probe completely from instrument and check functionality as described in the Setup
section on page 4 before assuming an instrument problem instead of a probe problem. If another probe is
available, this can prove useful in tracking down the source of the problem.
Error Code Table
NUMBER ERROR TYPE DESCRIPTION
System Hardware Errors
E0.0 NON-RECOVER Stack ram test error
E0.1 NON-RECOVER Internal RAM test failed
E0.2 NON-RECOVER Internal ROM check sum failed
E0.3 NON-RECOVER Instrument error
E0.4 RECOVERABLE EEPROM check sum error
E0.5 NON-RECOVER RAM check sum error
E0.6 RECOVERABLE EEPROM busy too long
E0.7 RECOVERABLE EEPROM address write error
Exception Errors
E1.0 NON-RECOVER External interrupt 0. (INT0.)
E1.2 NON-RECOVER Transmitter interrupt (TXD)
E1.3 NON-RECOVER Timer Y interrupt (TYINT)
E1.4 NON-RECOVER Timer 2 interrupt (T2INT)
E1.5 NON-RECOVER External counter interrupt 0. (CNTR0.)
E1.6 NON-RECOVER Timer 1 interrupt (T1INT)
E1.7 NON-RECOVER Serial I/O
E1.8 NON-RECOVER Software break
A/D Converter Errors
E2.0 RECOVERABLE Low cal. resistor pulse width error
E2.1 RECOVERABLE Hi cal. resistor pulse width error
E2.2 RECOVERABLE PTB resistor pulse width error
E2.3 RECOVERABLE Probe pulse width error
E2.4 RECOVERABLE Could not determine cause