Welch Allyn, Inc.
2 SureTemp
®
Model 670/SureTemp 4
®
Model 675
SPECIFICATIONS
•
Case Dimensions
(nominal):
6.25 inches long x 3.7 inches wide x 2.4 inches deep.
•
Case Material:
ABS Plastic.
•
Weight (nominal):
10.5 ounces with batteries.
•
Input:
Welch Allyn Thermistor Probe.
•
Display range:
28.9
o
C to 42.2
o
C (84.0
o
F to 108.0
o
F).
•
Laboratory Accuracy:
± 0.2
o
F in the Monitor mode and in a water bath per Welch Allyn
document number 90565-000.
•
Clinical Accuracy:
Meets the proposed ASTM clinical test criteria.
•
Push buttons:
Backlight/Recall and Pulse Timer.
•
Selection Switches:
Normal/Monitor and F/C (No F/C Switch for Euro/German version).
•
Probes:
Interchangeable Oral (also used for Axillary) and Rectal.
•
Power Source:
Three “AA” Batteries.
•
Battery Operating Life:
Up to 6,000 temperature measurements (At 72
o
F ambient temperature
and 15% backlight use).
•
Display Type:
Liquid Crystal Display, 3½ digits plus special icons.
•
Operating
temperature:
16
o
C to 40
o
C (60.8
o
F-104
o
F) @ 15% to 95% RH non-condensing
per ASTM E1112-86.
•
Storage Temperature:
-20
o
C to 50
o
C (-4
o
F-120
o
F) @ 15% to 95% RH non-condensing per
ASTM E1112-86.
OPERATIONAL CHARACTERISTICS
INTRODUCTION
Note:
This manual describes both the Model 670 and the Model 675 thermometers. The bulk of the
discussion is equally applicable to both products. Where there are differences, it will be noted as to which
instrument the discussion applies.
The Welch Allyn Model 670 and 675 Thermometers are the next generation of thermistor based medical
grade thermometers providing the accuracy and ease of use of thermistor technology as in the well known
Model 600, with enhancements in speed and features. Model 670 provides a Normal mode oral temperature
in about 10 seconds. Model 675 provides this temperature in about 4 seconds. These are significantly faster
than the typical 30 to 40 second average time to temperature in earlier products. Several other features
have been added, including a temperature recall feature, probe type display, displayed error codes for more
information when operation is abnormal, and a biotech mode to allow further investigation of some internal
parameters without the need for disassembly.