pH 3310 Operation
59
ba75802e02 12/2008
4.7 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.7.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH The following settings for pH measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the <F1>/[Menu] key
shortly
.
Setting Default settings
Buffer AutoCalTEC
Cal. interval 7 d
Unit for slope mV/pH
Measured parameter pH
Resolution pH 0.001
Resolution mV 0.1
Asymmetry 0 mV
Slope -59.16 mV
Man. temperature 25 °C
One point calibration Off