CI62L+RTL SPECTROPHOTOMETER
42
Technical Specifications
Measurement Geometrics d/8°, DRS spectral engine, choice of optical aperture:
14 mm measurement area/20 mm target window
Receiver Blue-enhanced silicon photodiodes
Light Source Gas-filled tungsten lamp
Illuminant Type A, C, D50, D55, D65, F2, F7, F11, and F12
Standard Observer 2° and 10°
Spectral Range 400nm – 700nm
Spectral Interval 10nm – measured, 10nm – output
Measurement Range 0 to 200% reflectance
Measuring Time Approx. 2 seconds
Inter-Instrument Agreement CIE L*a*b*:
Avg. 0.20 ∆E*
ab
, based on avg. of 12 BCRA series II tiles
(specular component included)
Max. 0.40 ∆E*
ab
on any tile (specular component included)
Short-Term Repeatability .05 ∆E*
ab
on white ceramic (standard deviation)
Lamp Life Approx. 500,000 measurements
Power Supply Removable Li-ion battery pack;
7.4 VDC rated @ 2400 mAh
AC Adapter Requirements Input 100-240 VAC, 50/60Hz
Output 12 VDC @ 2.5 A
Charge Time Approx. 4 hours – 100% capacity
Measurements Per Charge 1,000 measurements typical
Data Interface USB 2.0
Operating Temperature Range 50° to 104°F (10° to 40°C) 85% relative humidity maximum
(non-condensing)
Storage Temperature Range -4° to 122°F (-20° to 50°C)
Dimensions 4.3”H (10.9 cm) 3.6”W (9.1 cm) 8.4”L (21.3 cm)
Weight 2.32 lbs. (1.05 kg)
Accessories Provided Calibration Standard, Manual,
AC Adapter, Carrying Case
Usage Indoor only
Altitude 2000m
Pollution Degree 2
Overvoltage Category II
Specifications and design subject to change without notice.