5.0.2 Advanced Entry Calibration
Advanced Entry Calibration is the alternate method for special-case media
types. The procedure has two parts:
1) You must list the sensor readings for the label and TOF values using
different algorithms.
2) From your list the best algorithm is chosen, new readings taken and those
values entered.
+
Advanced Entry Calibration will override all previous calibration
settings; this method should be used only when Standard Calibration
has failed.
To perform an Advanced Entry Calibration:
Step Operator Action Displayed Message Comment
1
(Media should be loaded
and the appropriate
Sensor Type selected.)
Turn ‘On’ the printer.
Wait until the
printer initializes
(about six
seconds).
2
Press the Key.
Raise the printhead
assembly.
You are entering
Menu Mode.
3
Press the Key
to enter the Media
Settings menu.
See Section 4.1.2
for menu layout.
4
Press the Key
and scroll to ‘Sensor
Calibration’.
Press the
Key to abort this
procedure.
5
Press the
Key.
You are entering
the calibration
submenu.