Multi 3630 IDS What to do if...
ba77170e03 10/2017 107
Obviously incorrect
measured values
15.3 Dissolved oxygen
Error message,
OFL
Error message,
Error
15.4 Conductivity
Error message,
OFL
Error message,
Error
Cause Remedy
IDS-pH sensor:
IDS-pH sensor unsuitable Use suitable IDS sensor
Temperature difference
between buffer and test sample
too great
Adjust temperature of buffer or sam-
ple solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Measured value outside the
measuring range
Select a suitable measuring medium
Cause Remedy
Sensor contaminated Clean the sensor
Measured temperature value
outside the operating conditions
(display of OFL/UFL instead of a
temperature value)
Keep to the temperature range for
the test sample
Defective sensor Replace the sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Measured value outside the
measuring range
Use suitable IDS conductivity sen-
sor
Cause Remedy
Sensor contaminated Clean the sensor and replace it if
necessary
Calibration solution not suitable Check the calibration solutions