pH value Multi 3630 IDS
40 ba77170e03 10/2017
Calibration record
(USB output)
5.3 Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is dis-
played in the measured value display. Here you can very clearly see whether
or not the current measured value is in the calibrated part of the measuring
range.
The following information is displayed:
- -30 ... +30 -62 ... -61
or
-56 ... -50
Clean the IDS sensor according to the
sensor operating manual
Error Error
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to chapter
15 W
HAT TO DO IF...
For pH IDS sensors, you can optionally enable a more finely
graded calibration evaluation (QSC) (see section 5.4).
Display Calibration
record
Zero point
[mV]
Slope [mV/
pH]
Multi 3630 IDS
Ser. no. 09250023
CALIBRATIONpH
Calibration date 15.03.2016 16:13:33
SenTix 940
Ser. no. B092500013
TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope -60.2 mV/pH
Asymmetry 4.0 mV
Sensor +++
etc...