EasyManua.ls Logo

Xylem wtw Multi 3630 IDS - Manual Calibration (Concal)

Xylem wtw Multi 3630 IDS
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pH value Multi 3630 IDS
34 ba77170e03 10/2017
5.2.4 Manual calibration (ConCal)
Make sure that in the sensor menu, Buffer menu, the
ConCal
buffer set is cor-
rectly selected (see section 11.1.1).
Use one to five buffer solutions in any order. The pH values of the buffer solu-
tions should differ by at least one pH unit.
19 Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
20 If necessary, use <M> to finish the calibration.
The calibration record is displayed.
or
Switch to calibration with the next buffer with <ENTER>.
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
If single-point calibration was set in the menu, the calibration pro-
cedure is automatically finished with the measurement of buffer
solution 1 and the calibration record is displayed.
1 In the measured value display, select the measured parameter pH or
mV with <M>.
2 Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).

Table of Contents