EasyManua.ls Logo

Xylem wtw Multi 3630 IDS - Page 44

Xylem wtw Multi 3630 IDS
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
pH value Multi 3630 IDS
44 ba77170e03 10/2017
You can carry out QSC control calibrations at greater intervals than normal cal-
ibrations.
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).
Carrying out a QSC
control calibration
1QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
1
01.02.2016 08:00
Info
1 Open the menu for measurement settings with <ENTER>.
2 In the QSC menu, select Control calibration with <><>.
The calibration display appears. AutoCal QSC-Kit is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC control calibration.
3 Follow the user guide.
The calibration is carried out like a normal three-point calibration. As
soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
control calibration.

Table of Contents

Related product manuals