Multi 3630 IDS Dissolved oxygen
ba77170e03 10/2017 55
Calibration
evaluation FDO
®
925
Calibration record
(USB output)
Display Calibration
record
Relative slope
+++ S = 0.94 ... 1.06
++ S = 0.92 ... 0.94 or
S = 1.06 ... 1.08
+ S = 0.90 ... 0.92 or
S = 1.08 ... 1.10
Error Error
S < 0.90 or
S > 1.10
Perform error elimination accord-
ing to chapter 15 W
HAT TO DO IF...
Multi 3630 IDS
Ser. no. 10139695
CALIBRATION Ox
Calibration date 15.03.2016 16:13:33
FDO 925
Ser. no. 10146858
SC-FDO 925 10158765
Relative slope 0.98
Sensor +++