Multi 3630 IDS Settings
ba77170e03 10/2017 89
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
Conductivity The following settings for conductivity measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
Turbidity The following settings for the turbidity measurement are reset to the default
condition with the Reset function:
Setting Default settings
Calibration interval 180 d
Check interval 60 d
Measured parameter D.O. concentration
Relative slope (S
Rel
)1.00
Salinity (value) 0.0
Salinity (function) Off
Setting Default settings
Calibration interval 150 d
Measured parameter
ϰ
Cell constant (c) Depending on the connected
measuring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set)
0.100 cm
-1
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of
the linear temperature compensa-
tion
2.000 %/K
TDS factor 1.00