7-16
IM AQ6370C-17EN
SYSTEM
Function Control Command Remarks
OPTICAL ALIGNMENT
:CALibration:ALIGn[:IMMediate]
BUILT-IN SOURCE
:CALibration:INTernal[:IMMediate]
AQ6370D-L0
(Without Light
source)
EXTERNAL LASER
:CALibration:EXTernal[:IMMediate]
AQ6370D-L0
(Without Light
source)
WL CALIBRATION
BUILT-IN SOURCE
:CALibration:WAVelength:INTernal
[:IMMediate]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
EXTERNAL LASER
****.***nm
:CALibration:WAVelength:EXTernal
:SOURce<wsp>LASer|0;
:CALibration:WAVelength:EXTernal
:WAVelength<wsp><NRf>[M]
EXTERNAL GAS CELL
****.***nm
:CALibration:WAVelength:EXTernal
:SOURce<wsp>GASCell|1;
:CALibration:WAVelength
:EXTernal: WAVelength<wsp><NRf>[M]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
EMISSION LINE
****.***nm
:CALibration:WAVelength:SOURce
<wsp>EMISsion|2;
:CALibration:WAVelength:EXTernal
:WAVelength<wsp><NRf>[M]
AQ6373B
WL SHIFT **.***nm
:SENSe:CORRection:WAVelength:SHI
Ft<wsp><NRf>[M]
LVL SHIFT ***.***dB
:SENSe:CORRection:LEVel:SHIFt
<wsp><NRf>[DB]
WL OFFSET TABLE
:CALibration:WAVelength:OFFSet:T
ABLe<wsp><integer>,<NRf>[DB]
AQ6370C/AQ6370D/
AQ6373/AQ6373B/
AQ6375
LVL OFFSET TABLE
:CALibration:POWer:OFFSet:TABLe
<wsp><integer>,<NRf>[DB]
AQ6370C/AQ6370D/
AQ6373/AQ6373B/
AQ6375
GRID EDITOR
200GHz SPACING
:SYSTem:GRID<wsp>200GHZ|4
AQ6370C/AQ6370D/
AQ6375/AQ6375B
100GHz SPACING
:SYSTem:GRID<wsp>100GHZ|3
AQ6370C/AQ6370D/
AQ6375/AQ6375B
50GHz SPACING
:SYSTem:GRID<wsp>50GHZ|2
AQ6370C/AQ6370D/
AQ6375/AQ6375B
25GHz SPACING
:SYSTem:GRID<wsp>25GHZ|1
AQ6370C/AQ6370D/
AQ6375/AQ6375B
12.5GHz SPACING
:SYSTem:GRID<wsp>12.5GHZ|0
AQ6370C/AQ6370D/
AQ6375/AQ6375B
CUSTOM
:SYSTem:GRID<wsp>CUSTom|5
AQ6370C/AQ6370D/
AQ6375/AQ6375B
START WL ****.****nm
:SYSTem:GRID:CUSTom:STARt<wsp>
<NRf>[M|HZ]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
STOP WL ****.****nm
:SYSTem:GRID:CUSTom:STOP<wsp>
<NRf>[M|HZ]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
SPACING ***.*GHz
:SYSTem:GRID:CUSTom:SPACing
<wsp> <NRf>[GHZ]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
VALUE EDIT
-
INSERT
:SYSTem:GRID:CUSTom:INSert<wsp>
<NRf>[M|HZ]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
DELETE
:SYSTem:GRID:CUSTom:DELete<wsp>
<integer>
AQ6370C/AQ6370D/
AQ6375/AQ6375B
REFERENCE
WAVELENGTH ****.****nm
:SYSTem:GRID:REFerence<wsp><NRf>
[HZ]
AQ6370C/AQ6370D/
AQ6375/AQ6375B
USER KEY DEFINE
-
Note
On the AQ6375/AQ6375B, the GRID EDITOR soft key is not available when in Frequency mode.
7.2 Table of Correspondence between Soft Keys and Remote Commands