4-12
<Toc> <Ind> <4. BT200 BASIC OPERATIONS >
IM 01C00A11-01E
(3) CHANGE SETUP DATA
To change setup data, select a parameter to change
from parameter panel and press [ENT] to call setup
panel. If you have set a security code with a utility, that
code must be entered.
PARAM
C10:TAG NO.
TABC-101
C20:PRESS UNIT
mmH2O
C21:LRV
0 mmH2O
DATA DIAG PRNT ESC
F0417-1.EPS
SET
C10:TAG NO.
TABC-101
TABC-101
CODE CAPS CLP ESC
2.Enter data ‘EJ-aH01.’
1
2
PARAMETER PANEL
Example 1
Change the tag number
Current: TABC-101 EJ-aH01
If you have set a security code,
you must enter the security
code. (See section (4))
SETUP PANEL
1.Select the parameter to
change with [ ], [ ], [<] or [>]
and press [ENT].
<
>
F0417-2.EPS
( E ) ( J ) ( – ) ( a )
[F3] NO
[F2] POFF
Cancel the setup
procedure.
(See Section 5.2)
3
( H ) ( 01 ) 
]
]
SET
C10:TAG NO.
TABC-101
EJ-a01
*PRINTER OFF
*F2:PRINTER ON
FEED PDFF NO
3.Press [ENT].
SET
C10:TAG NO.
TABC-101
“EJ-a01 ”
*PRINTER OFF
*F2:PRINTER ON
FEED POFF NO
4.The setup data flashes for
your verification. Press [ENT]
to accept new tag number.
4
[F3] NO
[F4] OK
Retry the data setup.
Return to the
parameter panel.
SET
C10:TAG NO.
EJ-a01
FEED NO OK
5.The setup procedure is
complete. The data on the
connected instrument is
rewritten.
5

Related product manuals