Setting the Measurement Mode
Setting the Measurement Mode
The measurement mode determines how the entire 
GX/GP system operates. The GX/GP measurement 
characteristics change depending on the measurement 
mode. The measurement mode must be set before 
reconfiguration and before specifying various settings.
By factory default, the measurement mode is set to 
Normal. When performing high-speed or dual interval 
measurement according to measurement conditions, you 
need to set the measurement mode to High speed or Dual 
interval.
1
 Press MENU.
2
  Tap the Browse tab.
3
 Tap Initialize Calibration.
4
 Tap Measuremet mode.
5
  Setting the Measurement Mode.
6
 Tap Execute.
A conrmation screen is displayed.
7
 Tap OK
Operation complete
Note
•  When the measurement mode is changed, the system 
restarts, and the following data is initialized. Set the 
measurement mode before reconfiguration and before 
specifying various settings.
Data subject to initialization
All internal data
All setting parameters including security settings but 
excluding communication settings
System configuration data
•  You cannot set the measurement mode when 
recording, computation, or control execution is in 
progress.
•  The measurement mode is not initialized during 
initialization.
•  If the advanced security function (/AS) or multi-batch 
function (/BT) is enabled (On), the measurement 
mode is fixed to Normal.
  When changing the measurement mode, disable the 
functions beforehand.
Limitations
Depending on the measurement mode, there is a limit 
to the number of measurement channels, the number of 
recording channels, and the supported modules.
For the specific limitations, see the limitations provided in 
the following general specifications.
•  GX/10/GX20 Paperless Recorder (panel mount type)
  General Specifications GS 04L51B01-01EN
•  GP10/GP20 Paperless Recorder (portable type)
  General Specifications GS 04L52B01-01EN
This section explains the measurement mode, which needs to be set properly according to the measurement conditions.