<1. Overview>
1-2
IM 11M10A01-01E
Operating temperature range: Sensor 0 to 250°C (within 5 cm of tip, the rest 0 to 70°C )
Converter 0 to 50°C
Measurementgasconditions: Atmosphericpressure,temperature0to70°C,gasow
up to 1 m/s, no corrosive gases like SO
2
(can’t be used
inatmospherewithammablegases,corrosivegases,
or silicone vapor.
Power consumption: up to 15 VA.
Power supply : 100 V AC ±10% 50/60 Hz.
Weight : Body approx. 600 g, sensor approx 150 g.
Accessories : Oxygen sensor assembly (1), nipple (1), bracket (2),
extension cable(1), instruction manual (1), warranty
CARD (1)