<1. Overview>
1-3
IM 11M10A01-01E
1.3 ModelandSuxCodes
• OX100 Current limited Zirconia Oxygen Analyzer
Model SuxCode Option code Specications
OX100 ............................ ...................... General purpose Oxygen analyzer
Application -1
-2
-3
......................
......................
......................
Oxygenanalyzer(Can’tuseanactivatedcarbonlter)*1
For (N
2
)reowfurnace(Specifyincaseofusinganactivated
carbonlter)*1
MonitoringfordeciencyofOxygenwithlowalam,Measuring
range 0-25%
Measuring Range 1 ...................... 0-1000 ppm/ 0-25% O
2
Power Supply 5 ...................... 100 V AC ±10%, 50/60 Hz
Sensor 1
N
......................
......................
Sensor assy. cable 3m + extension 2 m
No sensor assembly
Manual E
J
......................
......................
English
Japanese
L9811LA plastic resin sensor holder is provided as standard accessory.
*1Formeasurementintherangeof1500to4500ppm,usetheOX102.
•Samplingunit,calibrationkit
Part Number Specications
K9424GA Sampling unit
K9424JA Calibrationkit:Six1-literasks,1000ppmO
2
/N
2
gasoutletcock(withowadjuster)
K9424JB Calibrationkit:Six1-literasks,1000ppmO
2
/N
2
gasoutletcock(withoutowadjuster)
K9424JR Calibrationkit:Six1-literasks,10000ppmO
2
/N
2
gasoutletcock(withowadjuster)
K9424JS Calibrationkit:Six1-literasks,10000ppmO
2
/N
2
gasoutletcock(withoutowadjuster)
•Accessories
Part Number Specications
K9346WM Activatedcarbonelement,veperpack:forsamplingunit
K9424JC Gasoutletcock(withowadjuster):qty.one
K9424JD Gasoutletcock(withoutowadjuster):qty.one
K9424JP Standard gas cylinder (1000 ppm O
2
/N
2
): qty. six
K9424JQ Standard gas cylinder (10000 ppm O
2
/N
2
): qty. six
K9424BA Sensorassy.cable3mforOX100only
K9424CA Extensioncable,2m:forOX100only
B9877BA BracketformountingOX100,twoperset.
K9424DB Sensorholder(metal)forOX100orOX102
L9811LA Sensorholder(plastic)forOX100orOX102
K9424GQ Disklter
K9424WH Feltlter
4th Edition : Oct. 23, 2019-00