1. Access output
Remove cover by releasing 4 screws.
Access to commissioning menu
Select output function
Select OUTP
Confirm selection
2. Adjust low span value
Display will show *4 mA
Adjust value for low span
Select digit to adjust
Adjust digit
Confirm adjusted value
3. Adjust high span value
Display will show *20 mA
Adjust value for high span
Select digit to adjust
Adjust digit
Confirm adjusted value
Return to measurement
4-2. Output range adjustment
YES
ENT
>
^
ENT
MODE
>
^
NO
IM 12B6C2-E-H
12
ESCAPE TO MEASURE can be used at any stage to abort operation.
WARNING: If the HOLD function is activated the instrument returns with the question HOLD (flashing); answer YES or NO or
MODE again to return to measurement.