<6. ELECTRONIC TRANSMITTER (-E)>
6-7
IM 01R01B02-00E-E 12th edition October 01, 2014 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
6.2.4 Selection of temperature unit (F15)
The function F15 sets the unit of temperature indication.
The following indications can be set : degC (Celsius) or degF (Fahrenheit).
At the factory the display is set to degC indication.
The selection of the indication is as follows :
Description Selection Key Indication
Change to setting mode Enter
Display mode
F1-
Setting mode Enter
4 x
Enter
F11
F15
F15 -1
Selection degC
degF
Enter
Enter
F15
F15 -2
F15
Back to display mode
F1-
Display mode
Note: If you press “" instead of “Enter”, you can return from the selected point to the previous menu
without activating the displayed parameter.
6.2.3 Totalizer reset (F14)
Function F14 resets the totalizer to zero.
The counter reset is performed as follows:
Description Selection Key Indication
Change to setting mode Enter
Display mode
F1-
Setting mode Enter
3 x
Enter
F11
F14
F14 -1
Selection Reset Enter F14
Back to display mode
F1-
Display mode
Note: If you press “" instead of “Enter”, you can return from the selected point to the previous menu
without activating the displayed parameter.