1.25 Initializing and Calibrating the System
(Initialization,reconfiguration,touchscreen
calibration)
Initialize the settings or the internal memory, reconfigure the GX/GP, and calibrate the touch
screen.
For details on A/D calibration, see page 5-4 in section 5.1.3, “Performing A/D Calibration and
Adjusting the Input Accuracy”.
For details on the advanced security settings (when the advanced security function (/
AS option) is in use), see section 2.1, “Enabling the Advanced Security Function,” in the
Advanced Security Function (/AS) User’s Manual (IM 04L51B01-05EN).
For details on encryption and certificates, see page 1-209 in section 1.26, “Configuring
Key Creation, Certificate Management, and Encryption/Certificate (SSL communication,
PDF electronic signature) (Release number 2 and later)”.
For details on updating the firmware (Web application, I/O module, and expansion
module), see page 5-11 in section 5.1.6, “Updating the Firmware (Release number 2 and later)”.
For the multi batch settings (multi-batch function (/BT option)), see section 2.1.1, “Settings
for the Multi-batch Function,” in the Multi-batch Function (/BT) User’s Manual (IM 04L51B01-
03EN).
1.25.1 Initializing the Settings and the Internal Memory
Path
GX/GP: MENU key > Browse tab > Initialize Calibration > Menu Initialize
Description
Setup Item Selectable Range or Options Default Value
Initialize all Off, On Off
Initialize all
Set this to On to initialize all settings and return the GX/GP to its factory default condition.
However, the A/D calibration password is not initialized.
Setup Item Selectable Range or Options Default Value
Security settings
1
Off, On Off
Other than Security settings
1
Off, On Off
Communication(IP address)
1, 2
Off, On Off
Communication(Server settings)
1, 2
Off, On Off
Calibration correction settings
1, 2
Off, On Off
Instruments tags, etc.
1, 2
Off, On Off
1 You can set this when Initialize all is set to Off.
2 You can set this when Other than Security settings is set to On. Release number 3 and later.
Security settings
Set this to On to initialize the security settings.
Other than Security settings
Set this to On to initialize settings other than the security settings.
Communication(IPaddress)
Set this to On to initialize the IP address, DHCP, and DNS settings.
Communication(Serversettings)
Set this to On when you want to initialize the server list settings, server encryption settings,
and channel conversion setting of DARWIN compatible communication.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA SMARTDAC+ GX10 and is the answer not in the manual?

YOKOGAWA SMARTDAC+ GX10 Specifications

General IconGeneral
BrandYOKOGAWA
ModelSMARTDAC+ GX10
CategoryVoice Recorder
LanguageEnglish

Related product manuals