46 MicroPro
© gbo Medizintechnik AG Version 1.4
Guidance and manufacturer’s declaration – electromagnetic emissions
The equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment should assure that it is used in such an environment.
Electromagnetic environment – guidance
The equipment must emit electromagnetic energy in order to per-
form its intended function. Nearby electronic equipment may be
affected.
The equipment is suitable in all establishments including than do-
mestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions, IEC
61000-3-2
Voltage fluctuation/flicker emis-
sions, IEC 61000-3-3
Guidance and manufacturer’s declaration — electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment should assure that it is used in such an environment.
Electromagnetic environment –
guidance
Electrostatic discharge (ESD),
IEC61000-4-2
8 kV contact discharge
15 kV air discharge
8 kV contact discharge
15 kV air discharge
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humid-
ity should be at least 30 %.
Electrical fast transient/burst,
IEC 61000-4-4
2 kV for power supply
lines
1 kV for input/output
lines
2 kV for power supply
lines
1 kV for input/output
lines
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
1 kV outer conductor –
outer conductor
2 kV outer conductor -
ground
1 kV outer conductor –
outer conductor
2 kV outer conductor -
ground
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines,
IEC 61000-4-11
0 % U
for ½ cycle in 45° steps
(100 % dip)
0 % U
for 1 cycle
(100 % dip)
70 % U
for 25 or 30 cycle at 50 or
60 Hz
(30 % dip)
0 % U
for 250 or 300 cycle at 50
or 60 Hz
(100 % dip)
0 % U
for ½ cycle in 45° steps
(100% dip)
0 % U
for 1 cycle
(100 % dip)
70 % U
for 25 or 30 cycle at 50 or
60 Hz
(30 % dip)
0 % U
for 250 or 300 cycle at 50
or 60 Hz
(100 % dip)
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
If the user of the equipment requires
continued operation during power
mains interruptions, it is recommend-
ed that the equipment be powered
from an uninterruptible power supply
or a battery.
Note: U
is the a.c. mains voltage prior to application of the test level.