Guidance and manufacturer’s declaration — electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment should assure that it is used in such an environment.
Electromagnetic environment –
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the equipment, including ca-
bles, than the recommended separation
distance calculated from the equation ap-
plicable to the frequency of the
transmitter.
Conducted HF disturbances,
IEC 61000-4-6
3 V
eff
150 kHz to 80 MHz
6 V
eff
in ISM and amateur
radio bands
3 V
eff
150 kHz to 80 MHz
6 V
eff
in ISM and ama-
teur radio bands
Recommended separation distance
d=1,2P
Radiated HF disturbances,
IEC 61000-4-3
Recommended separation distance
d=1,2P
for 80 MHz to 800 MHz
d=2,3P
for 800 MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts accord-
ing to the transmitter manufacturer and d
is the recommended separation distance
in meters (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Comments on software
This device uses free/open software and software from third parties, which meet the conditions of the
GNU General Public License v2 and v3 (GPLv2/GPLv3), the GNU Lesser General Public License v3
(LPGLv3) and/or other copyright licenses, disclaimers and notes.