Specifications
B-2 www.zoll.com 9650-000278-01 Rev. B
Alarm Volume ~ 80 dBA @ 1 meter
Noise Level ~ 60 dBA when measured at 1 meter
Operating Voltages AC Supply: 100 to 240 VAC (50/60 Hz)
Caution Use only the AC/DC power supply
that ZOLL provides with the
device.
DC Supply: Nominal 12.5 to 28.0 VDC
(accepts DC voltages between 11.8 to 30
VDC).
Caution Use only the DC power supply
cable that ZOLL provides with the
device.
Operating Time Internal Battery 3 hours using settings defined in
ISO 10079-1:2015 for battery operation
testing.
Aspirator Temperature Ranges
Operating Extreme: - 26
C to 55C
(- 15F to 131F)
Storage Temperature: - 46
C to 71C
(- 50F to 160F)
Battery Temperature Ranges
Battery Charge: 0
C to 45C (32F to 131F)
Battery Operation: - 26
C to 55C
(-15F to 131F)
Size 8.0” Wide x 12.5” High x 4.5” Deep
(20.3 cm Wide x 31.8 cm High x 11.4 cm
Deep)
Dimensions exclude accessories.
Weight
~9.9 lbs (4.5 kg)
Warranty Limited, 1 year