9650-000278-01 Rev. B ZOLL 330 Multifunction Aspirator Service Manual B-1
Appendix B
Specifications
General
Operation Modes Continuous/Intermittent
Applied Vacuum 10 mm Hg to 550 mm Hg
±(2mmHg+8%ofsetting)
Continuous Vacuum Range:
Surgical
Pharyngeal
Tracheal
CASS
Adjustable 20 to 550 mm Hg
Adjustable 60 to 400 mm Hg
Adjustable 60 to 220 mm Hg
Adjustable 10 to 40 mm Hg
Intermittent Vacuum Range ON/OFF Times Vacuum Level: Adjustable 10 to 200 mm Hg
ON/OFF Times: 5 to 40 seconds.
FREE AIRFLOW Continuous: Up to 30 LPM
Maximum FREE AIRFLOW: Up to 35 LPM*
*Maximum FREE AIRFLOW rate achieved
with the use of Smart Flow.
Collection Canister Capacity 1200 cc
Collection Cannister Material Polystyrene and durable polyethylene
LED Status/Alarm Indicator Red, Yellow, and Green

Table of Contents

Other manuals for ZOLL 330