Specifications
B-4 www.zoll.com 9650-000278-01 Rev. B
Emissions and Immunity Compliance
The ZOLL Aspirator is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The device design addresses EMC performance. Be sure to use the proper PWB identified in
this manual and make sure the proper power supply is supplied with this product.
The design does not rely on EMC gaskets or additional shielding to achieve compliance.
The operation manual identifies all applicable warnings for use in relation to Emissions and
Immunity Compliance.
Compliance for each emissions and immunity standard or test specified:
Emissions Test Compliance Electromagnetic Environment Guidance
The ZOLL aspirators are intended for use in the electromagnetic environment specified below. The
customer or user of the aspirator should ensure that they are used in such an environment.
RF emissions
CISPR 11
Group 1 The ZOLL aspirators use RF energy only for their internal
function. Therefore, their RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The ZOLL aspirators are suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEF 61000-3-2
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Immunity Test Test Level Compliance Level
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 8 kV contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 kV air
Electrical fast
transient/
burst
IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for
input/output lines
+/- 2 kV for power supply lines
+/- 1 kV for
input/output lines
Surge
IEC 61000-4-5
+/- 1 kV line(s) to
line(s)
+/- 2 kV line(s) to earth
+/- 1 kV line(s) to
line(s)
+/- 2 kV line(s) to earth
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
IEC 61000-4-11
0% Ut, 0.5 cycles at 0, 45, 90,
135, 180, 225, 270, and 315
degrees; 0% Ut, 1 cycle and
70% Ut, 25/30 cycles; single
phase at 0 degrees
0% Ut, 0.5 cycles at 0, 45, 90,
135, 180, 225, 270, and 315
degrees; 0% Ut, 1 cycle and
70% Ut, 25/30 cycles; single
phase at 0 degrees