906-0731-01-05 Rev. C ZOLL Ventilator Operator’s Guide 3-9
Mask CPAP Default Parameter Values
The Mask CPAP default parameter values are as follows:
Mask CPAP Parameter Start-Up Defaults
To select the unit’s default parameter values, highlight one of the above settings in the Start
Menu and press the ACCEPT/SELECT button. To operate with parameter values that differ
from the default values, use the unit’s Parameter buttons (see the “Changing Parameter Values”
section later in this chapter).
Note: You can configure the ventilator to automatically select Adult parameter defaults at
start up.
To adjust the parameter settings, always work from the lowest parameter button, Mode, to the
highest button, HR. The values that you select in a lower parameter window may affect the
values that appear in higher parameter windows
Warning! Never use the Noninvasive Positive Pressure Ventilation (NPPV) modes on a patient
that is NOT spontaneously breathing and/or may stop spontaneously breathing. CPAP
and BL are intended for ventilatory support, NOT ventilation.
When an NPPV mode is in operation, the head with mask icon appears in the location used by
the speaker/mute icons. Low and Medium priority alarms cause this head with mask icon to
disappear.
It reappears when Low priority alarms are muted.
When Medium priority alarms are muted, the muted speaker icon appears.
Mode CPAP
Backup BPM 12
Backup I:E 1:3.0
Backup PIP 20
PEEP 4
PIP Limit 30
FIO2 21